Techmark Precision Instrument Co.,Ltd.
  • Home
  • About
  • Event
  • Hot Disk
  • Thermtest
  • Knowledge
  • Solutions
    • FAQ
    • Support
  • English
    • Applications
圖片

TLS-100


Measures thermal conductivity and thermal resistivity. Transient line source. Portable thermal conductivity meter for testing soil, rock, concrete, and polymers.

​ASTM D5334 • ASTM D5930 • IEEE 442-1981

  • features
  • Specifications
<
>
The TLS-100 is a portable thermal conductivity meter used to measure thermal conductivity and thermal resistivity. Perfect for testing the thermal properties of soil, polymers, and other soft materials; as well as testing the thermal conductivity of concrete, rock, stone, or other hard materials. Tests can be performed with the push of a button. The collected data is automatically analyzed and results are displayed immediately.

The Transient Line Source (TLS) follows ASTM D5334. The sensor needle consists of a thin heating wire and temperature sensor sealed in a 100 or 50 mm steel tube. The sensor is completely inserted into the sample to be tested. Heat is delivered to the sample using a constant current source (q) and the temperature rise is recorded over a defined period of time. The slope (a) from a plot of temperature rise versus the logarithm of time is used in the calculation of thermal conductivity (k). The higher the thermal conductivity of a sample, the lower the slope. For samples of low thermal conductivity, the higher the slope.
​
  • Soil thermal conductivity testing concrete, polymers, soil and soft rock.
  • Follows international standards: ASTM D5334, ASTM D5930, & IEEE 442-1981
  • Portable, Economical, and Accurate
  • Easy to use
  • Standard 100 mm sensor for soft materials
  • Optional 50 mm sensor for hard materials
  • Soil Thermal Conductivity Meter
Materials
0.03 to 40 W/m/K.
Measurement Capabilities
0.1 to 20  mm²/s.
Thermal Conductivity
Up to 5 MJ/m³K.
Thermal Resistivity
2.5 to 160 seconds.
Measurement Time
Typically better than 3 %.
Reproducibility
Better than 5 % (thermal conductivity).
Accuracy
10 °C to 40 °C.
Temperature Range
Adjusted to the line voltage in the country of use.
​Smallest Sample (100 mm)
10 mm × 40 mm diameter or square.
Smallest Sample (50 mm)
Kapton sensor 8563 (9.9 mm radius)
Largest Sample Size
Unlimited
Standards
ASTM D5334, ASTM D5930, & IEEE 442-1981